Product : Plasma Sterilizers (Low Temperature Plasma Sterilizers)
Manufacturer : Healthwell Medical, Inc.
Manufacturer Country : KOREA
Details
1. Easy to use
You can start/cancel/control the sterilization process by simply pressing the key buttons. The selected
sterilization cycle will be progressed automatically and you can monito relevant information such as
sterilization conditions, progress status, error messages from the display.
2. Automatic notification system for regular maintenance
This product automatically notifies the replacement period for consumables such as pump oil, oil mist
trap and air filter. It displays an error message to replace the consumable once it reaches 250 cycles.
If the consumable is not replaced until it reaches 300 cycles, it will stop operation to prevent further
equipment failure.
3. Built-in self-test
When a malfunction occurs during operation, an error code is displayed by built-in self-test and
the running cycle is terminated automatically for safety.
4. Validation of Sterilization of HPS-30,50
A PCD(Process Challenge Device) in accordance with EU standards(EN 867-5) is used for
the validation of sterilization of a dead-end lumen at the most difficult positions inside of
the sterilization load.
-Test mode : Special Mode(Lumen Mode)
-Lumen Size : HPS-30,50 (1∅×1,000mm(L), Teflon)
Specifications
Sterilizing Agent : Hydrogen Peroxide, 1Cycle/Cartridge
Total Cycle Time : Smart Mode : 40min
Standard Mode : 48min
Special Mode : 56min
Cycle Temperature : 45℃~60℃
SAL(Sterility Assurance Level) : 10⎺⁶ (Bacillus)
By-products : Oxygen and vapor water only
Chamber : Shape : Rectangular
Material : Aluminum
Volume : Total : 30L
Dimensions : Overall : 594×525×605(W×H×D)mm
Chamber : 340×240×395(W×H×D)mm
Weight : 90kg
Control : 32bit Micro Processor
Cycle Information : 2.7” (OLED) Display, Printer
Electrical : AC 220-230V~, 50/60Hz, Max 2.7kw
Installation Requirements : Front : 100cm
Rear, Left side, Right side : 5cm
Placement : Built-in wheels provide mobility(Option Cart)
Room Conditions : Use Environment : 10~40℃, 30~75%RH
Installation Environment : -20~60℃, 0~95%RH
Printer : Built-in Thermal Printer(Brief Mode / Detail Mode)
Cycle Parameters(Temp, Pressure, Time, Daily&Total Cycle, etc)
Cart(Option) : 613×665×850(W×H×D)mm
* Total cycle time may vary depending on the conditions of the sterilization load such as
instrument type, quantity, dryness and so on.
Consumables
● Sterilant Cartridges
(Hydrogen Peroxide 50%)
1 cartridge/cycle
● Packing
Tyvek Pouch
Non-Woven
SheetImpulse Sealer
● Maintenance*
Vacuum Oil
Oil Mist Trap
*Recommend to be changed every 250 cycles. However, the number of cycles can vary according to daily practices of users in terms of sterilization loads(types, quantity, moisture content) and frequency of sterilization and etc.
● Sterilization Monitoring
CI(Chemical Indicator) : to monitor if the sterilization loads are exposed to H2O2 by a color change.
BI(Biological Indicator) : to monitor if the sterilization is successful by an incubation of bacillus.
● Exclusive Cart(Option)HPS-30 : 613×665×850(W×H×D)mm